Thermo Scientific iCAP TQ ICP-MS is a Triple Quadrupole Inductively Coupled Plasma Mass Spectrometer used for elemental determination at parts per trillion (ppt) levels. The sample in the liquid form is nebulized and transported into a high energy plasma where it gets ionised into its constituent ions. The ions generated are guided into quadrupole mass filter and quantified based on mass to charge ratio. Concentration is proportional to intensity of ions generated per second (cps).
       
Major Specifications:
        · Detection Limit (ng.L-1): 9Be <0.5, 115In <0.1 & 209Bi <0.1
        · Modes of Operation: SQ, SQKED & TQ
        · Sample Introduction Kits: Aqueous, HF, Organic & AGD (for High TDS)
        · RF Generator: 400 - 1600 W, 27 MHz
        · Plasma: Argon, Vacuum: TMP & Roughing Pumps (10-7 mbar)
        · Nebuliser: Micromist Quartz / PFA, Spray Chamber: Cyclonic Quartz / PFA
        · Torch: Quartz, Injector: Quartz / Sapphire
        · Sampler & Skimmer Cones: Nickel & Platinum, Ion Optics: 90° Lens
        · Q1 Quadrupole: 4 MHz, 2-240 u, User Definable Resolution
        · Q2 Quadrupole: Collision Reaction Cell (CRC) with MFC for He, H2 & O2
        · Q3 Quadrupole: 2 MHz, 2-290 u, Scan Speed > 3700 u/s, Li to U with 40 Interval Masses, User Definable Resolution
        · Abundance Sensitivity in TQ Mode: < 0.005 ppm
        · Ion Detection: Dynode Electron Multiplier, Simultaneous Pulse/Analog, Dwell Time 100µs, Dynamic Range 1010 cps
        · Software: Qtegra
        · Thermofisher iSC 65 Autosampler: 240 Positions
        · Applications: Semi-Quantitative, Quantitative, Isotope Ratio, Isotope Dilution, Nano Particle Separation and Counting. Seamless Integration with Ion Chromatograph (for Fe, Cr, Hg & As Speciation) and Laser Ablation (for Geo Applications). Extended Applications Include Drug Discovery.
        · Sample Required: 500 mg – 1 g (Solid) & 10 ml (Liquid)
 
 
 
 

 

 

 
   
   
 