Gas Chromatograph - Mass Spectrometer (GC-MS)

 
Description:
Agilent Model 8890 GC System with Single Quadrupole Mass Spectrometer (5977B MSD) analyzer is used for the separation and identification of thermally stable volatile compounds. The GC consists of Split / Splitless (SSL) injectors and capillary columns for different applications. NIST spectral library search is also available.
 
Major Specifications:
        · GC oven temperature: Near ambient -450°C
        · Chromatographic performance:
            Area repeatability: <0.5%
            Retention time repeatability <0.008%
        · Inlet Split ratio: 7500:1
        · Mass filter: Heated monolithic hyperbolic quadrupole
        · Ionization modes: Electron Impact(EI) and Chemical Ionization(CI)
        · Mass Range: (m/z) 1.6 −1,050 amu
        · Mass resolution: Unit mass
        · Mass Accuracy: 1 μL injection of 100 a pg/μL OFN standard scanning from 50.350 u will give its monoisotope at m/z 271 ±0.005
        · Spectral Accuracy: 1 μL injection of a 100 pg/μL OFN standard scanning from 50.350 u will give 99.0% spectral accuracy
        · Ion source temperature: 150 -350°C
        · Quadrupole Temperature: 106-200°C
        · Detector Triple-Axis Detector with high energy dynode and long life electron multiplier
 
Accessories/Attachments:
        · Auto liquid sampler with 50 vial capacity
        · Capillary Columns: Polar Columns (DB-WAX) & HP-5 MS UI
        · Spectral Library and Database:: licensed NIST 2017 Library
        · Software : Open Lab CDS 2.5 version
 
Sample Requirements:
        · Powder samples: 5-10mg
        · Liquid samples: 0.2-1ml
 
Applications:
        · Quality Control
        · Mixture Analysis
        · Structural Analysis of Natural Products, Petrochemicals & Pharmaceuticals
        · Impurity Profiling
        · Metabolomics
        · Nominal molecular weight calculations
 

 

 

 
   
   
 